Model No: WM-IP02

Infusion Pump

Quantity :
Contact Now
 
WM-IP02 Infusion Pump Features:
 
1. Infusion Pump is with the classic vertical design. The software compensation error is within±5% after calibration. It is widely used in ICU, NICU, pediatrics, internal medicine, surgery, obstetrics and gynecology, oncology, emergency department and outpatient infusion room.
2. Double CPU and numeric keypad make it safe and more user friendly.
3. Adjustable Levels Setting: 10 levels setting for occlusion, air bubble and alarm volume
4. Strong Battery Backup: 3000mAh battery and backup time more than 10hrs.
5. User Friendly: Big Screen and numeric keypad
6. Dual Occlusion Sensor: Up and down occlusion sensor with DPS system
7. Infusion Mode: Volume mode, time mode

WM-IP02 Infusion Pump Technical Specification:

Infusion Mode
Volume Mode, Time Mode, Drug Mode
Rate Unit
ml/h, volume/time, d/min
IV Sets
Standard single use IV sets ( 10,15,20,60d/mL, with diameter 3.4 ~4.5mm)
Flow Rate
1.0ml/h~1200ml/h 1.0d/min~400d/min
Increment
1.0-99.9ml/h in 0.1ml/h step;100-1200ml/ in 1ml/h step
1-400d/min in 1d/min step;
Deviation
Flow Rate:≤5%(after calibration)
VTBI
0.1~9999.9 ml
Total Volume
0.0~9999.9 ml
Time
1min~99h00min
Purge Rate
600ml/h,display purge volume
KVO
1.0ml/h~5.0ml/h,adjustable
Keypad
Numeric keypad,
BOLUS
Bolus function available, flow rate adjustable
Night Mode
Screen to be dim automatically in 5mins after standby time expired.
Occlusion pressure detector
downstream pressure detection, 40KPa~130KPa,
10 levels adjustable with dynamic  pressure value display
Air Bubble Detector
Ultrasonic detection, ≥50ul,Single bubble volume: 50~300 μl,
10 levels adjustable.
Volume
10 levels adjustable
Motor Anti-reverse Function
Anti-reverse function to prevent upstream
IV Set Management
Open system, accept different IV brands after calibration
Self-Testing
Self-Testing before startup and during infusion
Audio Visual Alarm Message
Air bubble, downstream occlusion, door open, infusion finish, standby time expired, battery low, battery empty, battery disengaged, AC Power disengaged, motor error, pump error, pressure sensor error, parameter error, system error
Screen
3.2inch brightness LCD
Power Supply
AC Power:100V~240V,50Hz/60Hz,≤25VA
Battery
Rechargeable Lithium battery, DC11.1V,3000mAh;
operating time: ≥10h @25ml/h, displaying remaining battery
Operation Condition
Temperature:+5℃~+40℃, Humidity:20%~90%,
Air pressure:70~106KPa
Transport & Storage
Temperature :-20℃~+55℃,Humidity:10%~95%,
Air pressure:50~106KPa
Classification
Type CF, Class I, IPX3
  Dimension
145mm×150mm×200mm;Weight:1.7kg